基于纳米技术的血红蛋白生物传感器快速测定油炸食品的丙烯酰胺
CSTR:
作者:
作者单位:

作者简介:

作者简介:费永乐(1989-),女,硕士研究生,研究方向:食品科学与安全 通讯作者:李书国(1969-),男,博士,教授,研究方向为粮油食品科学与安全技术

通讯作者:

中图分类号:

基金项目:

国家自然科学基金面上项目(20876165);河北省教育厅科技项目(2009329)


Rapid Detection of Acrylamide in Fried Food by Nanotechnology-based Hemoglobin Biosensor
Author:
Affiliation:

Fund Project:

  • 摘要
  • |
  • 图/表
  • |
  • 访问统计
  • |
  • 参考文献
  • |
  • 相似文献
  • |
  • 引证文献
  • |
  • 资源附件
  • |
  • 文章评论
    摘要:

    丙烯酰胺是一种神经毒素,具有潜在致癌性。丙烯酰胺分子可以和血红蛋白结构中N-末端缬氨酸的α-NH2形成共价化合物引起电极钝化,其含量不同,钝化程度也不同。基于此原理,本研究将多壁碳纳米管/血红蛋白/壳聚糖修饰的生物传感器用于油炸食品中丙烯酰胺的检测。利用差分伏安脉冲法对油炸食品中丙烯酰胺进行定量分析并对条件优化,多壁碳纳米管修饰量为10 μg/cm2、支持电解质溶液为0.1 mol/L的PBS溶液(pH=7.4,含5×10-3 mol/L K4[Fe(CN)6],0.1 mol/L NaCl),其中K4[Fe(CN)6]作为氧化还原探针,电位增量0.008 V、脉冲幅度0.05 V、脉冲宽度0.1 V、脉冲间隔0.1 s,丙烯酰胺线性检测范围3.0×10-8 mol/L~3.0×10-7 mol/L,最低检测限(S/N=3)为1.2×10-8 mol/L。与高效液相色谱法相比,该法具有简便、快速、准确、样品预处理简单、无需衍生化等优点。

    Abstract:

    Acrylamide is a neurotoxin and a potential carcinogen. Hemoglobin (Hb) biosensor passivation can be achieved by the formation of a covalent compound by acrylamide with the -NH2 group of the N-terminal valine of hemoglobin. Different concentrations of acrylamide produce different degrees of passivation. Based on this principle, hemoglobin biosensors were prepared by immobilizing multi-wall carbon nanotubes (MWCNTs) and Hb in a chitosan film on the surface of glassy carbon electrodes and these were used for the detection of acrylamide in fried food. The optimum analysis conditions for the detection of acrylamide in fried food by differential pulse voltammetry (DPV) was determined as follows: pulse increment 0.008 V, pulse amplitude 0.05 V, pulse width 0.1 V, pulse interval 0.1 s. The oxidative-reductive detector consisted of 10 μg/cm2 MWCNTs and the supporting electrolytes were 0.1 mol/L PBS (pH 7.4), 0.1 mol/L NaCl and 5×10-3 mol/L potassium ferrocyanide. The linear range of response was from 3.0×10-8 mol/L to 3.0×10-7 mol/L and the detection limit was estimated as 1.2×10-8 mol/L, taking into account a signal-to-noise ratio of 3. The developed method is practical in the determination of acrylamide in fried foods, with the advantages of simplicity, rapidity, sensitivity, and accuracy over the standard high-pressure liquid chromatography method.

    参考文献
    相似文献
    引证文献
引用本文

费永乐,王丽然,李书国.基于纳米技术的血红蛋白生物传感器快速测定油炸食品的丙烯酰胺[J].现代食品科技,2015,31(2):268-273.

复制
相关视频

分享
文章指标
  • 点击次数:
  • 下载次数:
  • HTML阅读次数:
  • 引用次数:
历史
  • 收稿日期:2014-07-03
  • 最后修改日期:
  • 录用日期:
  • 在线发布日期: 2015-02-10
  • 出版日期:
文章二维码
×
因办公室装修,期间暂时无法接听电话,如有事请QQ或邮件联系。信息咨询:QQ: 2553003667稿件处理1:QQ: 1542354573稿件处理2:QQ: 2195608851 财务咨询:QQ: 1347040116 Email:mfood@scut.edu.cn、mfood@foxmail.com